Seres2000型氨氮操作说明一、引泵二、校准循环1 21/09/10 09:58 Remotestopped..Menu=>C1 21/09/10 09:58 Remotestopped..Menu=>CS1 amo 00,00mg/LS1 amo 00,00mg/L按“C”键按“C”键UserMenu=>A ExtensionSERES=>BUserMenu=>A ExtensionSERES=>BEscape=>EEscape=>E按“A”键进入用户菜单按“A”键进入用户菜单1Senddates 2Local/remoteCTRL1Senddates 2Local/remoteCTRL3calibrationcoef 4Modifycoef3calibrationcoef 4Modifycoef5Restoreconfig 6ResetAlarms5Restoreconfig 6ResetAlarmsEscape=>ENext=>DReturn=>BEscape=>ENext=>DReturn=>B按“D”键下翻至第三页按“2”键进入校准循环菜单1Enterdate/time2ModifycycledelayCalibrationcoefficient3Pumpspriming4Option5Option6ZeromeasuringValidate=>CEscape=>EEscape=>ENext=>DReturn=>B按“C”键下一步按“3”键进入引泵菜单Calibrationonline=>1Withstandard=>2PumpPriming(1,2,3,4,5,6)state=0PushrightkeyforON/OFFValidate=>CEscape=>EEscape=>E按“C”键下一步按数字键1-6分别引动相对应的泵SelectStandardpositionPumpPriming(1,2,3,4,5,6)123456on"Sample/Standard"switchPushrightkeyforON/OFFDipthetubeintothestandardsolutionEscape=>ENext=>D按数字键“E”退出引泵按“D”键开始校准循环校准循环完成后注意事项:、泵3和泵5不能同时引动,因硫酸和水杨酸钠Adjustindicatedvaluewithkey会生成絮状物水杨酸,致测量室堵塞;key2=>+key3=>—Validatewithkey02、引泵要充分,要确保管路当中没有气泡存在。按数字键“2”,“3”加减校准值按数字键“0”退出校准循环注意事项:1、通过修改数字键盘上的“2”“3”键,使屏幕右上方显示的值和校准液一致,按“0”键(约3秒)确认修改成功后,仪器将完成余下的清洗循环。2、仪器每次校准循环后会自动做一次测量循环,如不需要,可跳过,继续做校准循环。3、校准一般要做2—3次,直至仪器稳定。Seres2000型氨氮操作说明一、引泵二、校准循环1 21/09/10 09:58 Remotestopped..Menu=>C1 21/09/10 09:58 Remotestopped..Menu=>CS1 amo 00,00mg/LS1 amo 00,00mg/L按
COD说明书 来自淘豆网m.daumloan.com转载请标明出处.